Surface Cleaning and Disinfection

of Anesthesia Devices


The cleaning and disinfection of anesthesia device surfaces are essential for infection prevention

and are a critical process to ensure patient safety.

Cleaning


The initial step removes organic matter, inorganic debris, and contaminants from equipment surfaces, reducing conditions for microbial growth.


This step improves the effectiveness of disinfectants and is performed mechanically using soap or detergents at different pH levels.


Afterward, residues must be thoroughly rinsed off and the equipment fully dried.


Disinfection


A process performed after cleaning to reduce harmful microorganisms on equipment surfaces.


Chemical or thermal methods are chosen based on device features and disinfectant type, using only validated approaches.


Registered products must be used exactly as specified in the manufacturer’s instructions.



Sterilization


A procedure that thoroughly kills all microorganisms, including spores.


It is most effective when performed after the removal of organic material and pathogens.


Common methods include physical techniques (e.g., steam) and chemical processes (e.g., ethylene oxide gas).




Anesthesia Machine After-Use Cleaning and Disinfection in Infectious Patients


1. Pre-entry risk assessment

Assess equipment disconnection risks based on patient condition.


2. Hand hygiene

Perform handwashing or hand disinfection before entering the room or before cleaning.


3. Personal protective equipment (PPE)

Wear appropriate PPE, including a mask, eye protection, gloves, a waterproof gown, and protective footwear.


4. Power shutdown and patient disconnection

Disconnect the device from the patient and turn off the power or remove the battery.


5. Relocation to a well-ventilated area

Relocate equipment to an area with proper airflow.


6. Cleaning with detergent

Wipe all surfaces, sensors, and cables with detergent and a cloth or disposable wipe, then dry with a clean cloth.

Avoid using too much liquid.


7. Disinfection

Apply disinfectant thoroughly from top to bottom in a well-ventilated area.

Avoid electrical connections, sensors, and cable interfaces.


8. Disposal

Immediately dispose of all disposable materials.